Blumenthal EZ, Frenkel S.

Eye. 2005 Mar;19(3):308-11.
Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. eblumenthal@md.huji.ac.il

ABSTRACT:

OBJECTIVE: To evaluate the interdevice reproducibility of retinal nerve fibre layer (RNFL) thickness measurements obtained with the commercially available GDx-VCC, a scanning laser polarimeter with variable (individualized) corneal compensation.

METHODS: A prospective instrument validation study in which 13 GDx-VCC devices were tested. One eye each, from three normal subjects were used to test each of the devices, on the same day, by an experienced operator. Variability and reproducibility for each of five GDx parameters were calculated.

RESULTS: For each of five tested GDx parameters, the coefficient of variation and 95% confidence interval range (microm), for the 13 devices, respectively, were: TSNIT avg: 5.1%, 3.84 microm; Superior avg: 5.3%, 4.82 microm; Inferior avg: 6.1%, 5.50 microm; TSNIT standard deviation: 8.6%, 2.92 microm; and nerve fibre indicator (NFI): N/A, 5.69. Item reliability (Cronbach’s alpha) for the five GDx parameters are: TSNIT-Avg: 0.97, Sup-Avg: 1.00, Inf-Avg: 0.84, TSNIT-SD: 0.99, NFI: 0.99.

CONCLUSIONS: With the commercially available GDx-VCC, our results indicate that RNFL measurements appear reproducible across devices.

Article as Full-Text PDF

Related Posts

Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
List of Publications- by Prof. Eytan Blumenthal
ORIGINAL RESEARCH PAPERS Blumenthal EZ. Adjusting the slit-lamp oculars: An unnecessary burden or a must?  Survey of Ophthalmology 1995;40(3):225-228. Abstract Full-text (PDF) Blumenthal EZ, Serpetopoulos...
Adjusting The Slit-Lamp Oculars: An Unnecessary Burden Or A Must?
Blumenthal EZ. Surv Ophthalmol. 1995 Nov-Dec;40(3):225-8. Department of Opthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjusting the slit-lamp eyepiece rings is perceived by some...
On Focusing The Slit-Lamp: Part I. An Inaccurate Ocular Setting--What Is There To Lose?
Blumenthal EZ, Serpetopoulos CN. Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjustments of the diopter rings on the...
Early Transient Visual Acuity Loss After Lasik Due To Steroid-Induced Elevation Of Intraocular Pressure
Frucht-Pery J, Landau D, Raiskup F, Orucov F, Strassman E, Blumenthal EZ, Solomon A. J Refract Surg. 2007 Mar;23(3):244-51. Department of Ophthalmology, Hadassah-Hebrew University Medical Center, Jerusalem,...