Sanchez-Galeana C, Bowd C, Blumenthal EZ, Gokhale PA, Zangwill LM, Weinreb RN.

Ophthalmology. 2001 Oct;108(10):1812-8.
Glaucoma Center, University of California San Diego, La Jolla, California 92093-0946, USA.

ABSTRACT:

PURPOSE: To evaluate the sensitivity and specificity for discriminating between early to moderate glaucomatous and normal eyes using summary data reports from the Heidelberg Retina Tomograph (HRT), the GDx Nerve Fiber Analyzer (GDx), and the Optical Coherence Tomograph (OCT).

DESIGN: Comparative cross-sectional study.

PARTICIPANTS: One eye each of 50 normal subjects and 39 glaucoma patients with early to moderate visual field damage (mean deviation, -5.04 +/- 3.32 dB; range, -0.85 to -13.2 dB).

METHODS: Three experienced graders masked to patient identity and diagnosis evaluated each summary data report from the HRT, GDx, and OCT independently.

MAIN OUTCOME MEASURES: Each summary report was classified as either normal or glaucomatous. Sensitivity and specificity are reported for each grader, and agreement between graders is reported.

RESULTS: For the HRT, sensitivity and specificity ranged from 64% to 75% and 68% to 80%, respectively. Agreement (kappa +/- standard error [SE]) between observers one and two, two and three, and one and three was 0.73 +/- 0.07, 0.77 +/- 0.07, and 0.67 +/- 0.08, respectively. For the GDx, sensitivity and specificity ranged from 72% to 82% and 56% to 82%, respectively. Agreement (kappa +/- SE) between observers one and two, two and three, and one and three was 0.66 +/- 0.08, 0.66 +/- 0.08, and 0.50 +/- 0.09, respectively. For the OCT, sensitivity and specificity ranged from 76% to 79% and 68% to 81%, respectively. Agreement (kappa +/- SE) between observers one and two, two and three, and one and three was 0.73 +/- 0.07, 0.58 +/- 0.08, and 0.51 +/- 0.09, respectively.

CONCLUSIONS: When used alone, HRT, GDx, and OCT summary data reports can differentiate between normal and glaucomatous eyes with mild to moderate visual field loss. However, none of the instruments provided sensitivity and specificity that justify summary data reports being used as a screening tool for early to moderate glaucoma.

Article as Full-Text PDF

Figure from this article:

Related Posts

Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
List of Publications- by Prof. Eytan Blumenthal
ORIGINAL RESEARCH PAPERS Blumenthal EZ. Adjusting the slit-lamp oculars: An unnecessary burden or a must?  Survey of Ophthalmology 1995;40(3):225-228. Abstract Full-text (PDF) Blumenthal EZ, Serpetopoulos...
Adjusting The Slit-Lamp Oculars: An Unnecessary Burden Or A Must?
Blumenthal EZ. Surv Ophthalmol. 1995 Nov-Dec;40(3):225-8. Department of Opthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjusting the slit-lamp eyepiece rings is perceived by some...
On Focusing The Slit-Lamp: Part I. An Inaccurate Ocular Setting--What Is There To Lose?
Blumenthal EZ, Serpetopoulos CN. Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjustments of the diopter rings on the...
Early Transient Visual Acuity Loss After Lasik Due To Steroid-Induced Elevation Of Intraocular Pressure
Frucht-Pery J, Landau D, Raiskup F, Orucov F, Strassman E, Blumenthal EZ, Solomon A. J Refract Surg. 2007 Mar;23(3):244-51. Department of Ophthalmology, Hadassah-Hebrew University Medical Center, Jerusalem,...