Frenkel S, Slonim E, Horani A, Molcho M, Barzel I, Blumenthal EZ.

Ophthalmology. 2005 Feb;112(2):257-61.
Department of Ophthalmology, Hebrew University-Hadassah Medical Center, Jerusalem, Israel.

ABSTRACT:

PURPOSE: To ascertain operator learning effect, and to quantify the interoperator reproducibility, using the newly introduced GDx VCC (variable corneal compensator) scanning laser polarimeter.

DESIGN: Prospective instrument validation study.

PARTICIPANTS: Three operators with no prior experience in operating the GDx VCC examined one randomly chosen eye of each of 30 randomly ordered subjects (15 glaucoma patients and 15 normal subjects).

METHODS: Each study eye was scanned by the 3 operators in a random sequence during a single session. Five GDx parameters were analyzed: TSNIT (temporal, superior, nasal, inferior, temporal) Average, TSNIT Standard Deviation (SD), Superior Average, Inferior Average, and Nerve Fiber Indicator (NFI).

MAIN OUTCOME MEASURES: Retinal nerve fiber layer (RNFL) thickness GDx parameters across operators.

RESULTS: A learning effect was not found for any of the operators for any of the 5 GDx parameters studied. Reproducibility for each GDx parameter was measured as the mean SD value of the measurements taken by the 3 operators. Interoperator reproducibilities for normal and glaucomatous eyes, respectively, were 0.8 and 1.6 microm (TSNIT Average), 1.7 and 2.2 microm (Superior Average), 1.6 and 2.7 microm (Inferior Average), 1.3 and 1.7 microm (TSNIT SD), and 1.4 and 4.6 microm (NFI). The data obtained by each operator were highly correlated with those of the other 2 operators.

CONCLUSIONS: With the commercially available GDx VCC, a learning effect was not found for 3 novice operators. In addition, RNFL measurements seem highly reproducible across operators.

Article as Full-Text PDF

Related Posts

Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
List of Publications- by Prof. Eytan Blumenthal
ORIGINAL RESEARCH PAPERS Blumenthal EZ. Adjusting the slit-lamp oculars: An unnecessary burden or a must?  Survey of Ophthalmology 1995;40(3):225-228. Abstract Full-text (PDF) Blumenthal EZ, Serpetopoulos...
Adjusting The Slit-Lamp Oculars: An Unnecessary Burden Or A Must?
Blumenthal EZ. Surv Ophthalmol. 1995 Nov-Dec;40(3):225-8. Department of Opthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjusting the slit-lamp eyepiece rings is perceived by some...
On Focusing The Slit-Lamp: Part I. An Inaccurate Ocular Setting--What Is There To Lose?
Blumenthal EZ, Serpetopoulos CN. Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjustments of the diopter rings on the...
Early Transient Visual Acuity Loss After Lasik Due To Steroid-Induced Elevation Of Intraocular Pressure
Frucht-Pery J, Landau D, Raiskup F, Orucov F, Strassman E, Blumenthal EZ, Solomon A. J Refract Surg. 2007 Mar;23(3):244-51. Department of Ophthalmology, Hadassah-Hebrew University Medical Center, Jerusalem,...