Blumenthal EZ, Serpetopoulos CN.

Ophthalmology. 2000 Feb;107(2):329-33.
Department of Ophthalmology, Hadassah University Hospital, and Hebrew University-Hadassah Medical School, Jerusalem, Israel.

ABSTRACT:

PURPOSE: Meaningful errors in photocoagulation spot size may result from several factors. In this article we discuss one major factor, namely, fluctuations in the surgeon’s accommodative state, coupled with an inaccurate setting of the slit-lamp oculars.

DESIGN: Experimental study.

METHODS: We analyzed and tested the optics of slit-lamp mounted lasers. Varying the ocular setting is correlated with measurements of the actual spot size obtained with each system.

MAIN OUTCOME MEASURE: The spot size obtained.

RESULTS: Three distinct, but related, phenomena that may lead to spot size errors are defined: (1) focusing the laser spot as opposed to focusing the retinal image; (2) instrument misalignment; (3) inadvertent accommodation.

CONCLUSION: The ocular setting must be meticulously calibrated to produce a true spot-sized burn. At the 50 microm setting, each diopter of induced accommodation, or erroneous ocular setting, almost doubles the actual spot size obtained. With large (500 microm) spot size settings, the defocused delivery system is more prone to spot-size errors in contrast with parfocal lasers.

Article as Full-Text PDF

Figures from this article:

Related Posts

Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
List of Publications- by Prof. Eytan Blumenthal
ORIGINAL RESEARCH PAPERS Blumenthal EZ. Adjusting the slit-lamp oculars: An unnecessary burden or a must?  Survey of Ophthalmology 1995;40(3):225-228. Abstract Full-text (PDF) Blumenthal EZ, Serpetopoulos...
Adjusting The Slit-Lamp Oculars: An Unnecessary Burden Or A Must?
Blumenthal EZ. Surv Ophthalmol. 1995 Nov-Dec;40(3):225-8. Department of Opthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjusting the slit-lamp eyepiece rings is perceived by some...
On Focusing The Slit-Lamp: Part I. An Inaccurate Ocular Setting--What Is There To Lose?
Blumenthal EZ, Serpetopoulos CN. Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjustments of the diopter rings on the...
Early Transient Visual Acuity Loss After Lasik Due To Steroid-Induced Elevation Of Intraocular Pressure
Frucht-Pery J, Landau D, Raiskup F, Orucov F, Strassman E, Blumenthal EZ, Solomon A. J Refract Surg. 2007 Mar;23(3):244-51. Department of Ophthalmology, Hadassah-Hebrew University Medical Center, Jerusalem,...